Characterization and Test of a Data Acquisition System for PET

Morrocchi, Matteo, Marcatili, Sara, Belcari, Nicola, Bisogni, Maria Giuseppina, Collazuol, Gianmaria, Ambrosi, Giovanni, Corsi, Francesco, Foresta, Maurizio, Marzocca, Cristoforo, Matarrese, Gianvito, Sportelli, Giancarlo and Guerra Gutiérrez, Pedro (2011). Characterization and Test of a Data Acquisition System for PET. In: "2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)", 23/10/2011 - 29/10/2011, Valencia, España. ISBN 978-1-4673-0118-3.

Description

Title: Characterization and Test of a Data Acquisition System for PET
Author/s:
  • Morrocchi, Matteo
  • Marcatili, Sara
  • Belcari, Nicola
  • Bisogni, Maria Giuseppina
  • Collazuol, Gianmaria
  • Ambrosi, Giovanni
  • Corsi, Francesco
  • Foresta, Maurizio
  • Marzocca, Cristoforo
  • Matarrese, Gianvito
  • Sportelli, Giancarlo
  • Guerra Gutiérrez, Pedro
Item Type: Presentation at Congress or Conference (Article)
Event Title: 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
Event Dates: 23/10/2011 - 29/10/2011
Event Location: Valencia, España
Title of Book: Proceedings of 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
Date: 2011
ISBN: 978-1-4673-0118-3
Subjects:
Faculty: E.T.S.I. Telecomunicación (UPM)
Department: Ingeniería Electrónica
Creative Commons Licenses: Recognition - No derivative works - Non commercial

Full text

[thumbnail of INVE_MEM_2011_104615.pdf]
Preview
PDF - Requires a PDF viewer, such as GSview, Xpdf or Adobe Acrobat Reader
Download (491kB) | Preview

Abstract

A small Positron Emission Tomography demonstrator based on LYSO slabs and Silicon Photomultiplier matrices is under construction at the University and INFN of Pisa. In this paper we present the characterization results of the read-out electronics and of the detection system. Two SiPM matrices, composed by 8 × 8 SiPM pixels, 1.5 mm pitch, have been coupled one to one to a LYSO crystals array. Custom Front-End ASICs were used to read the 64 channels of each matrix. Data from each Front-End were multiplexed and sent to a DAQ board for the digital conversion; a motherboard collects the data and communicates with a host computer through a USB port. Specific tests were carried out on the system in order to assess its performance. Futhermore we have measured some of the most important parameters of the system for PET application.

More information

Item ID: 12048
DC Identifier: https://oa.upm.es/12048/
OAI Identifier: oai:oa.upm.es:12048
Official URL: http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumb...
Deposited by: Memoria Investigacion
Deposited on: 17 Sep 2012 09:41
Last Modified: 21 Apr 2016 11:12
  • Logo InvestigaM (UPM)
  • Logo GEOUP4
  • Logo Open Access
  • Open Access
  • Logo Sherpa/Romeo
    Check whether the anglo-saxon journal in which you have published an article allows you to also publish it under open access.
  • Logo Dulcinea
    Check whether the spanish journal in which you have published an article allows you to also publish it under open access.
  • Logo de Recolecta
  • Logo del Observatorio I+D+i UPM
  • Logo de OpenCourseWare UPM