Validation of a small-animal PET simulation using GAMOS: a Geant4-based framework

Cañadas Castro, Mario, Arce, Pedro and Rato Mendes, Pedro (2011). Validation of a small-animal PET simulation using GAMOS: a Geant4-based framework. "Physics in Medicine and Biology", v. 56 (n. 1); pp. 273-288. ISSN 0031-9155. https://doi.org/10.1088/0031-9155/56/1/016.

Description

Title: Validation of a small-animal PET simulation using GAMOS: a Geant4-based framework
Author/s:
  • Cañadas Castro, Mario
  • Arce, Pedro
  • Rato Mendes, Pedro
Item Type: Article
Título de Revista/Publicación: Physics in Medicine and Biology
Date: January 2011
ISSN: 0031-9155
Volume: 56
Subjects:
Faculty: E.T.S.I. Telecomunicación (UPM)
Department: Otro
Creative Commons Licenses: Recognition - No derivative works - Non commercial

Full text

[thumbnail of INVE_MEM_2011_77000.pdf]
Preview
PDF - Requires a PDF viewer, such as GSview, Xpdf or Adobe Acrobat Reader
Download (987kB) | Preview

Abstract

onte Carlo-based modelling is a powerful tool to help in the design and optimization of positron emission tomography (PET) systems. The performance of these systems depends on several parameters, such as detector physical characteristics, shielding or electronics, whose effects can be studied on the basis of realistic simulated data. The aim of this paper is to validate a comprehensive study of the Raytest ClearPET small-animal PET scanner using a new Monte Carlo simulation platform which has been developed at CIEMAT (Madrid, Spain), called GAMOS (GEANT4-based Architecture for Medicine-Oriented Simulations). This toolkit, based on the GEANT4 code, was originally designed to cover multiple applications in the field of medical physics from radiotherapy to nuclear medicine, but has since been applied by some of its users in other fields of physics, such as neutron shielding, space physics, high energy physics, etc. Our simulation model includes the relevant characteristics of the ClearPET system, namely, the double layer of scintillator crystals in phoswich configuration, the rotating gantry, the presence of intrinsic radioactivity in the crystals or the storage of single events for an off-line coincidence sorting. Simulated results are contrasted with experimental acquisitions including studies of spatial resolution, sensitivity, scatter fraction and count rates in accordance with the National Electrical Manufacturers Association (NEMA) NU 4-2008 protocol. Spatial resolution results showed a discrepancy between simulated and measured values equal to 8.4% (with a maximum FWHM difference over all measurement directions of 0.5 mm). Sensitivity results differ less than 1% for a 250–750 keV energy window. Simulated and measured count rates agree well within a wide range of activities, including under electronic saturation of the system (the measured peak of total coincidences, for the mouse-sized phantom, was 250.8 kcps reached at 0.95 MBq mL−1 and the simulated peak was 247.1 kcps at 0.87 MBq mL−1). Agreement better than 3% was obtained in the scatter fraction comparison study. We also measured and simulated a mini-Derenzo phantom obtaining images with similar quality using iterative reconstruction methods. We concluded that the overall performance of the simulation showed good agreement with the measured results and validates the GAMOS package for PET applications. Furthermore, its ease of use and flexibility recommends it as an excellent tool to optimize design features or image reconstruction techniques.

More information

Item ID: 7082
DC Identifier: https://oa.upm.es/7082/
OAI Identifier: oai:oa.upm.es:7082
DOI: 10.1088/0031-9155/56/1/016
Official URL: http://iopscience.iop.org/0031-9155/56/1/016
Deposited by: Memoria Investigacion
Deposited on: 25 May 2011 09:06
Last Modified: 20 Apr 2016 16:12
  • Logo InvestigaM (UPM)
  • Logo GEOUP4
  • Logo Open Access
  • Open Access
  • Logo Sherpa/Romeo
    Check whether the anglo-saxon journal in which you have published an article allows you to also publish it under open access.
  • Logo Dulcinea
    Check whether the spanish journal in which you have published an article allows you to also publish it under open access.
  • Logo de Recolecta
  • Logo del Observatorio I+D+i UPM
  • Logo de OpenCourseWare UPM