Texto completo
|
PDF (Portable Document Format)
- Acceso permitido solamente a usuarios en el campus de la UPM
- Se necesita un visor de ficheros PDF, como GSview, Xpdf o Adobe Acrobat Reader
Descargar (1MB) |
ORCID: https://orcid.org/0000-0002-2986-7228 and Manuel Martín, Fernando Jaime
ORCID: https://orcid.org/0000-0003-4225-6744
(2013).
Wave reflection at a stent.
"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine", v. 227
(n. 1);
pp. 72-77.
ISSN 0954-4119.
https://doi.org/10.1177/0954411912456785.
| Título: | Wave reflection at a stent |
|---|---|
| Autor/es: |
|
| Tipo de Documento: | Artículo |
| Título de Revista/Publicación: | Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine |
| Fecha: | Enero 2013 |
| ISSN: | 0954-4119 |
| Volumen: | 227 |
| Número: | 1 |
| Materias: | |
| ODS: | |
| Palabras Clave Informales: | Stent, hemodynamics, fluid mechanics, wave reflections |
| Escuela: | E.T.S.I. Industriales (UPM) |
| Departamento: | Ingeniería Energética y Fluidomecánica [hasta 2014] |
| Licencias Creative Commons: | Reconocimiento - Sin obra derivada - No comercial |
|
PDF (Portable Document Format)
- Acceso permitido solamente a usuarios en el campus de la UPM
- Se necesita un visor de ficheros PDF, como GSview, Xpdf o Adobe Acrobat Reader
Descargar (1MB) |
A simple analytical expression has been derived to calculate the characteristics of a wave that reflects at a stent implanted in a uniform vessel. The stent is characterized by its length and the wave velocity in the stented region. The reflected wave is proportional to the time derivative of the incident wave. The reflection coefficient is a small quantity of the order of the length of the stent divided by the wavelength of the unstented vessel. The results obtained coincide with those obtained numerically by Charonko et al. The main simplifications used are small amplitude of the waves so that equations can be linearized and that the length of the stent is small enough so that the values of the wave functions are nearly uniform along the stent. Both assumptions hold in typical situations.
| ID de Registro: | 15402 |
|---|---|
| Identificador DC: | https://oa.upm.es/15402/ |
| Identificador OAI: | oai:oa.upm.es:15402 |
| URL Portal Científico: | https://portalcientifico.upm.es/es/ipublic/item/571942 |
| Identificador DOI: | 10.1177/0954411912456785 |
| URL Oficial: | http://pih.sagepub.com/content/227/1/72.abstract |
| Depositado por: | Memoria Investigacion |
| Depositado el: | 28 Abr 2015 15:27 |
| Ultima Modificación: | 12 Nov 2025 00:00 |
Publicar en el Archivo Digital desde el Portal Científico