A Novel Active Device for Shoulder Rotation Based on Force Control

Alguacil Diego, Isabel M. ORCID: https://orcid.org/0000-0002-2054-3622, Cuesta Gómez, Alicia ORCID: https://orcid.org/0000-0001-9507-2717, Pont Esteban, David ORCID: https://orcid.org/0000-0001-5889-7816, Carrillo Ríos, Juan, Espinosa Peralta, Paul ORCID: https://orcid.org/0000-0001-9290-1081, Sánchez Urán, Miguel Ángel ORCID: https://orcid.org/0000-0001-6652-0090 and Ferre Pérez, Manuel ORCID: https://orcid.org/0000-0003-0030-1551 (2023). A Novel Active Device for Shoulder Rotation Based on Force Control. "Sensors", v. 23 (n. 13); p. 6158. ISSN 1424-8220. https://doi.org/10.3390/s23136158.

Descripción

Título: A Novel Active Device for Shoulder Rotation Based on Force Control
Autor/es:
Tipo de Documento: Artículo
Título de Revista/Publicación: Sensors
Fecha: 5 Julio 2023
ISSN: 1424-8220
Volumen: 23
Número: 13
Materias:
Palabras Clave Informales: haptic; medical rehabilitation; rehabilitation; shoulder movements; upper limb; Exercise; Exercise Therapy; Force control; Haptic; Humans; medical rehabilitation; Range of Motion, Articular; Rehabilitation; Shoulder; Shoulder Joint; shoulder movements; Upper Extremity; Upper Limb
Escuela: E.T.S.I. Industriales (UPM)
Departamento: Automática, Ingeniería Eléctrica y Electrónica e Informática Industrial
Licencias Creative Commons: Reconocimiento - Sin obra derivada - No comercial

Texto completo

[thumbnail of 10090636.pdf] PDF (Portable Document Format) - Se necesita un visor de ficheros PDF, como GSview, Xpdf o Adobe Acrobat Reader
Descargar (3MB)

Resumen

This article describes a one-degree-of-freedom haptic device that can be applied to perform three different exercises for shoulder rehabilitation. The device is based on a force control architecture and an adaptive speed PI controller. It is a portable equipment that is easy to use for any patient, and was optimized for rehabilitating external rotation movements of the shoulder in patients in whom this was limited by muscle-skeletal injuries. The sample consisted of 12 shoulder rehabilitation sessions with different shoulder pathologies that limited their range of shoulder mobility. The mean and standard deviations of the external rotation of shoulder were 42.91 ± 4.53° for the pre-intervention measurements and 53.88 ± 4.26° for the post-intervention measurement. In addition, patients reported high levels of acceptance of the device. Scores on the SUS questionnaire ranged from 65 to 97.5, with an average score of 82.70 ± 9.21, indicating a high degree of acceptance. The preliminary results suggest that the use of this device and the incorporation of such equipment into rehabilitation services could be of great help for patients in their rehabilitation process and for physiotherapists in applying their therapies.

Más información

ID de Registro: 84959
Identificador DC: https://oa.upm.es/84959/
Identificador OAI: oai:oa.upm.es:84959
URL Portal Científico: https://portalcientifico.upm.es/es/ipublic/item/10090636
Identificador DOI: 10.3390/s23136158
URL Oficial: https://www.mdpi.com/1424-8220/23/13/6158
Depositado por: iMarina Portal Científico
Depositado el: 21 Nov 2024 15:32
Ultima Modificación: 21 Nov 2024 16:56