Texto completo
Vista Previa |
PDF (Portable Document Format)
- Se necesita un visor de ficheros PDF, como GSview, Xpdf o Adobe Acrobat Reader
Descargar (312kB) | Vista Previa |
ORCID: https://orcid.org/0000-0002-4871-8682, Martín Caro, Carlos, Pigozzi, Fabio, Calderón Montero, Francisco Javier, Alvarez, María, Maffulli, Nicola, Benito Peinado, Pedro José
ORCID: https://orcid.org/0000-0002-1631-8182, Di Salvo, Valter and Díaz Molina, Víctor
(2008).
Validation of a new portable metabolic system during an incremental running test.
"Journal of Sports Science and Medicine", v. 7
(n. 4);
pp. 525-531.
ISSN 1303-2968.
| Título: | Validation of a new portable metabolic system during an incremental running test |
|---|---|
| Autor/es: |
|
| Tipo de Documento: | Artículo |
| Título de Revista/Publicación: | Journal of Sports Science and Medicine |
| Fecha: | Diciembre 2008 |
| ISSN: | 1303-2968 |
| Volumen: | 7 |
| Número: | 4 |
| Materias: | |
| ODS: | |
| Palabras Clave Informales: | accuracy,consistency,oxygen consumption,portable metabolic chart,running metabolism ,running ,exercise ,regression analysis,physical fitness |
| Escuela: | Facultad de Ciencias de la Actividad Física y del Deporte (INEF) (UPM) |
| Departamento: | Otro |
| Licencias Creative Commons: | Reconocimiento - Sin obra derivada - No comercial |
Vista Previa |
PDF (Portable Document Format)
- Se necesita un visor de ficheros PDF, como GSview, Xpdf o Adobe Acrobat Reader
Descargar (312kB) | Vista Previa |
We tested a new portable metabolic system, the Jaeger Oxycon Mobile (OM) at a range of running speeds. Six subjects carried out, in random order, two incremental tests on a treadmill, one of them using the OM, and the other using the Jaeger Oxycon Pro (OP). There are systematic errors in the measurements of oxygen consumption (VO2) and respiratory exchange ratio (RER) with the OM. Production of CO2 (VCO2) tends to be overestimated by the OM, although the differences are not significant. Ventilation (VE) showed very similar values in both analyzers. Data of VO2 and RER were corrected with a regression equation which minimised the differences among the devices. The portable metabolic system OM makes systematic errors in measurements of VO2 and RER which can be adjusted with a regression analysis to obtain data comparable to those obtained by fixed systems
| ID de Registro: | 2184 |
|---|---|
| Identificador DC: | https://oa.upm.es/2184/ |
| Identificador OAI: | oai:oa.upm.es:2184 |
| URL Oficial: | http://www.jssm.org/b-v7n4.php |
| Depositado por: | Memoria Investigacion |
| Depositado el: | 08 Feb 2010 12:47 |
| Ultima Modificación: | 19 Jun 2024 08:12 |
Publicar en el Archivo Digital desde el Portal Científico